Calibrating the FCR25 Probe Changer

The calibration process for the Renishaw FCR25 Probe Changer represents the steps that you would take to prepare to use any of the probe changers that PC-DMIS supports. The process discussed here is specific to the FCR25 Probe Changer.

Two inserts are used with the FCR25 rack to accommodate different probe types: PA25-SH and PA25-20.

PA25-SH and PA25-20 Inserts

PA25-SH and PA25-20 Inserts

For example, the following image of an FCR25 Probe Changer shows three ports: two with inserts and one without (probes are also shown). The left port shows a PA25-SH insert, and the right port shows a PA25-20 insert. The middle port does not have an insert.

FCR25 Probe Changer inserts and probe types

The FCR25 Probe Changer must be mounted on the machine table parallel to either the X or Y axis for successful calibration. The FCR25 Probe Changer can be mounted to your table with the MRS rack or the three and six port stand-alone racks. For mounting instructions, consult the documentation that came with your probe changer.

To calibrate your probe changer:

Step 1 - Select the FCR25 Probe Changer

Step 2 - Define the Mount Point

Step 3 - Define the Ports

Step 4 - Prepare the Probe Changer

Step 5 - Begin Calibration

Step 6 - Measure Port 1 / PA25-SH Insert

Step 7 - Measure Port 3 / PA25-20 Insert

Step 8 - Measure Port 2 / No Insert

Step 9 - Review the Calibration Results