Scanning

Requirements

In order to use PC-DMIS to scan your parts with an analog probe (SP600 or SP25), there are several requirements that need to be met.

Functionality

The chart below summarizes the available PC-DMIS scanning features and the status of their support on Sheffield CMM’s with the SMP400.

 

PC-DMIS SCAN TYPES SUPPORTED? NOTES
LINEAROPEN YES ONLY LINE TECHNIQUE (CONSTANT INCREMENT) AND SPHERICAL BOUNDARY ARE SUPPORTED ON SMP400.
LINEARCLOSED YES ONLY LINE TECHNIQUE (CONSTANT INCREMENT) AND SPHERICAL BOUNDARY ARE SUPPORTED ON SMP400.
PATCH YES ONLY LINE TECHNIQUE (CONSTANT INCREMENT) AND SPHERE BOUNDARY ARE SUPPORTED ON SMP400.
CIRCLE YES OPEN-LOOP AND CLOSED
CYLINDER YES OPEN-LOOP AND CLOSED
AXIS YES CLOSED LOOP
CENTER YES CLOSED LOOP
LINE YES CLOSED LOOP
     
ROTARY NO FUTURE DEVELOPMENT EFFORT
SECTION NO TESTING PHASE. NOT RELEASED AT THIS TIME.
UV NO TESTING PHASE. NOT RELEASED AT THIS TIME. TOUCH MODE ONLY
FREEFORM NO DEFINED SCANS ARE NEEDED FOR IMPLEMENTATION, AND ARE NOT SUPPORTED ON SMP400.
PERIMETER NO NEEDS DEFINED SCANS THAT ARE NOT SUPPORTED ON SMP400.CLOSED LOOP

 

Unsupported scan controls/methods

Scanning Limitations

OPEN/CLOSED LOOP Scanning

If you wish to control the type of scan being performed for a measurement routine, you must quit PC-DMIS and launch the Settings Editor.

OPEN LOOP UseClosedLoop FALSE

CLOSED LOOP UseClosedLoop TRUE

There are no controls to change to OPEN or CLOSED by PC-DMIS commands, or by the loss of surface with an OPEN loop. The global value set in by the Settings Editor program is a global default.

Scan Speeds

Recommended Settings

Settings Editor

Settings Name Recommended Value Comment
MovePositionTol 1  
MPHits 1  
MPVectors 1  
PointDensity 4  
ScanSpeed 150 ScanSpeed is the systems? MAXIMUM allowable scanning speed.
SendProbeOffsets TRUE  
SMP400Baud 57600 or 115200 (recommended) Baud settings are configured on board the SMP400 by jumper pin location. Contact Hardware Technical Support regarding changing from 57600 to 115200.
UpdateComm FALSE  
UseVectorTouch FALSE  
UseNewCancel TRUE  
UsePartAlignment FALSE  

Parameter Settings - Opt. Probe tab

Settings Name Recommended Value Comment
# of Return Data   0 Set this to Zero To disable controller based scan filtering. PC-DMIS requires unfiltered scan data from the controller. Filtered data will result in an undersized measurement.
Positioning Accuracy 10  
Probing Accuracy 1  
PointDensity See comment This represents the number of readings /mm to record and is application specific to set to any particular number.
Offset Force CMM Model Dependent .150 for Endeavor CMMs .250 for Discovery and Apollo CMMs