To verify your rotary table calibration:
Select the Operation | CNC Programming | Create CNC Measurement Routine menu option to export the calibration measurement routine.
Open the NC Gateway server application.
Click Download Program, browse, and select the newly-created CNC measurement routine.
Click Analyze, and execute the calibration measurement routine on the NC machine.
Note the updated calibration information in the Rotary Table Setup dialog box in PC-DMIS after the calibration finishes. You may want to add dimensions for the calibration measurement routine to view the results of the calibration in a report.
Check the Calibration - Create a sphere measurement routine with the rotary table enabled. In this measurement routine, the sphere is always at zero degrees in PC-DMIS, but the location of each measurement deviates by the error in calibration of the table.