Active Tip List

(This item pertains to the Probe Utilities dialog box, accessible through Insert | Hardware Definition | Probe.)

PC-DMIS is capable of storing data that describes a large number of probe tips. This data includes the probe tip's ID, rotation, type, location, direction, diameter and thickness, date and time of calibration, and any non-calibrated tips. These are all stored inside the Active Tip List.

Active Tip List

There can be as many as 32767 probe tips stored. This number may be limited by the amount of disk space available on your system.

PC-DMIS describes a probe according to the following criteria:


Tip List Description

1. TIP ID#
2. Tip Rotation
3. Tip Type
4. X,Y,Z Location
5. I,J,K Direction
6. Diameter and Thickness
7. Date and Time

* (asterisk) - Noncalibrated Tip

Adding Tips to the List

You can define new tips, adding them to the list, by using the Add Angles button. See the "Add Angles" topic.

Editing Tip Data

  1. Highlight the desired active tip in the Active Tip List.

  2. Press the Edit button.

An Edit dialog box appears, allowing you to change the displayed values.

Non-calibrated tips have an asterisk before their Tip ID # in the Active Tip List.

Setting the Calibration Order

The calibration order is determined by the order in which the tips are chosen from the list box.

To set the calibration order:

  1. Select the User Defined option in the Measure Probe dialog box (See the "Measure" topic).

  2. Hold down the Ctrl key on the keyboard.

  3. With your left mouse button, select the tips from the Active Tip List for calibration. A number representing the index of the tip measurement order appears next to each Tip ID as it is selected.

If no tips are chosen, PC-DMIS prompts if you want to measure all tips.

Selecting a Tip to Use

You can define a specific probe tip to use in your measurement routine in these ways:

The Edit window command line for a sample tip would read:
TIP/T1A0B0, SHANKIJK=0, 0, 1, ANGLE=0

Until PC-DMIS encounters another TIP command in the measurement routine flow, it uses that tip.

Deleting a Tip or Wrist Angle

To delete one or more as tip from the Active Tip List:

  1. Select the item or items to be removed.

  2. Click the Delete button.

PC-DMIS also allows specific wrist angles to be removed from the Active Tip List. If a tip or probe head is deleted, all related wrist angles are also deleted from the file.

This option deletes the tip from the Active Tip List and removes it from the file of tips known to the system.