Registry Entry Optiv Classic 321 Manual Optiv Classic 321 GL Optiv Classic 453 All Other Options Description
Home_SpeedRestrict

TRUE

TRUE

TRUE

TRUE

Restricts the homing speed. The software displays an error message if the machine exceeds the  maximum allowed speed.

AxisX_Resolution

-0.00005

-0.00005

-0.00005

-0.00005

Defines the X-axis encoder resolution in mm.

AxisX_HomeEnabled

 

TRUE

TRUE

TRUE

When set to TRUE, the X-axis homing is enabled and required.

AxisX_HomeVelocityFactor

 

0.25

0.05

0.25

Defines the X-axis homing velocity as a percentage of the maximum speed. The valid values are 0.0 - 1.0 inclusive (0% - 100%).

AxisX_HomeFollowingLimitFactor

 

0.25

0.1

0.25

Defines the X-axis homing following limit as a percentage of the maximum following limit. The valid values are 0.0 - 1.0 inclusive (0% -100%).

AxisY_Resolution

0.00005

0.00005

0.00005

0.00005

Defines the Y-axis encoder resolution in mm.

AxisY_HomeEnabled

 

TRUE

TRUE

TRUE

When set to TRUE, the Y-axis homing is enabled and required.

AxisY_HomeVelocityFactor

 

0.25

0.05

0.25

Defines the Y-axis homing velocity as a percentage  of the maximum speed. The valid values are 0.0 - 1.0 inclusive (0% -100%).

AxisY_HomeFollowingLimitFactor

 

0.25

0.1

0.25

Defines the Y-axis homing following limit as a percentage of the maximum following limit. The valid values are 0.0 - 1.0 inclusive (0% -100%).

AxisZ_Resolution

-0.00005

-0.00005

-0.00005

-0.00005

Defines the Z-axis encoder resolution in mm.

AxisZ_HomeEnabled

 

TRUE

TRUE

TRUE

When set to TRUE, the Z-axis homing is enabled and required.

AxisZ_HomeVelocityFactor

 

0.25

0.05

0.25

Defines the Z-axis homing velocity as a percentage of the maximum speed. The valid values are 0.0 - 1.0 inclusive (0% -100%).

AxisZ_HomeFollowingLimitFactor

 

0.25

0.1

0.25

Defines the Z-axis homing following limit as a percentage of the maximum following limit. The valid values are 0.0 - 1.0 inclusive (0% -100%).

AxisQ_Resolution

 

0.006

0.006

0.006

Defines the rotary Q-axis encoder resolution in degrees.

AxisQ_HomeEnabled

 

FALSE

FALSE

FALSE

When set to TRUE, the rotary Q-axis homing is enabled and required.

Zoom_Resolution

-0.001

-0.001

-0.001

-0.001

Defines the zoom axis encoder resolution.

Zoom_Velocity

6.0

 

 

 

Defines the zoom axis velocity.

Zoom_Acceleration

45.0

 

 

 

Defines the zoom axis acceleration.

Zoom_Jerk

0.0

 

 

 

Defines the zoom axis tuning parameter Jerk.

Zoom_PFactor

50.0

 

 

 

Defines the zoom axis tuning parameter Proportional Gain Factor.

Zoom_DFactor

0.0

 

 

 

Defines the zoom axis tuning parameter Derivative Gain Factor.

Zoom_IFactor

0.0

 

 

 

Defines the zoom axis tuning parameter Integral Gain Factor.

Zoom_ILimit

0.0

 

 

 

Defines the zoom axis tuning parameter Integral Limit.

Zoom_ContThreshold

10.0

 

 

 

Defines the zoom axis tuning parameter Continuation Threshold.

Zoom_FollowingLimit

0.05

 

 

 

Defines the zoom axis tuning parameter Following Error Limit.

Zoom_BacklashCorrection

500

500

500

500

Defines the zoom overshoot distance to counter Backlash (encoder counts) Correction.

Illumination_SettleTime

0.1

0.1

0.1

0.1

Defines the delay in seconds for the illumination to settle after a change.

Stage_SettleTime

 

0.1

0.1

0.1

Defines the delay in seconds for the stage movement to settle after the software detects a motion.

Zoom_SettleTime

0.1

0.1

0.1

0.1

Defines the delay in seconds for the optics to settle after a magnification change occurs.

Pendant_EStopInhibitsAmp

 

TRUE

TRUE

TRUE

When set to TRUE, the emergency stop (ESTOP) disables the drive amps.

Pendant_FineSpeedFactor

 

1.0

1.0

1.0

Defines the fine speed as a percentage  of the maximum speed. The valid values are 0.0 - 1.0 inclusive (0% - 100%).

Pendant_X_Reversed

 

 

 

FALSE

When set to TRUE, the X-axis pendant motion is reversed.

Pendant_X_Deadband

 

 

 

0.2

Defines the X-axis pendant motion deadband at the center position in percent.

Pendant_X_Speed

 

 

 

1.0

Defines the X-axis pendant movement speed.

Pendant_X_FineSpeed

 

 

 

0.1

Defines the X-axis pendant fine movement speed.

Pendant_X_Exponential

 

 

 

0.5

Defines the X-axis pendant motion curve.

Pendant_Y_Reversed

 

 

 

TRUE

When set to TRUE, the Y-axis pendant motion is reversed.

Pendant_Y_Deadband

 

 

 

0.2

Defines the Y-axis pendant motion deadband at the center position in percent.

Pendant_Y_Speed

 

 

 

1.0

Defines the Y-axis pendant movement speed.

Pendant_Y_FineSpeed

 

 

 

0.1

Defines the Y-axis pendant fine movement speed.

Pendant_Y_Exponential

 

 

 

0.5

Defines the Y-axis pendant motion curve.

Pendant_Z_Reversed

 

 

 

FALSE

When set to TRUE, the Z-axis pendant motion is reversed.

Pendant_Z_Deadband

 

 

 

0.1

Defines the Z-axis pendant motion deadband at the center position in percent.

Pendant_Z_Speed

 

 

 

1.0

Defines the Z-axis pendant movement speed.

Pendant_Z_FineSpeed

 

 

 

0.1

Defines the Z-axis pendant fine movement speed.

Pendant_Z_Exponential

 

 

 

0.5

Defines the Z-axis pendant motion curve.

Pendant_Q_Reversed

 

 

 

FALSE

When set to TRUE, the rotary Q-axis pendant motion is reversed.

Pendant_Q_Deadband

 

 

 

0.1

Defines the rotary Q-axis pendant motion deadband at the center position in percent.

Pendant_Q_Speed

 

 

 

1.0

Defines the rotary Q-axis pendant movement speed.

Pendant_Q_FineSpeed

 

 

 

0.1

Defines the rotary Q-axis pendant fine movement speed.

Pendant_Q_Exponential

 

 

 

0.5

Defines the rotary Q-axis pendant motion curve.

Probe_DebounceTime

 

0.05

0.05

0.05

Defines the time in seconds that the touch probe must be triggered to detect a hit.

Probe_DirectionThreshold

 

0.01

0.01

0.01

Defines the distance in mm that the touch probe must move to determine the hit vector.

Probe_DisableIfNotActive

 

TRUE

TRUE

TRUE

When set to TRUE, the software disables the touch probe if it is not the active probe.

Probe_UsePI200

 

FALSE

FALSE

FALSE

When set to TRUE, the software uses the Renishaw PI200/TP200 probe.