An attempt was made to calibrate a tip other than the master tip (tip #0) when the calibration fixture's location relative to tip 0 was unknown. Tip 0 must be calibrated before any other tip can be calibrated. Note that this error may occur if a tip 0 calibration operation was started and then aborted.
As a result, the command is not executed.
To fix this error, calibrate tip 0.