The amp/check/divide board detected a counter fault on this axis. If this axis uses a scale and reading head, then the possible causes are:
The gratings are dirty.
There is too much DC offset in the Amp Check Board Lissajous.
There is not sufficient amplitude in the read head signals in the Amp Check Board Lissajous.
The read head skew angle is improper, as viewed on the Amp Check Board Lissajous.
The axis is moving too fast (over 40 in/sec. for standard machines and over 25 in/sec. for laser machines or over 15 in/sec. on some Discovery machines).
If this axis uses a laser interferometer, then the possible causes are:
The laser beam is blocked.
The laser beam is not properly aligned in the interferometer.
There is too much DC offset in the OPD Lissajous.
There is not sufficient amplitude in the OPD Lissajous.
The axis is moving too fast (over 25 in/sec. for laser machines).
What you should do:
To clear the error, click Clear error in the XYZ Display window.
To clear it using an RCU-II, press and hold down the Servo Power Off button and then press the Joy button twice.
If the axis uses a scale and reading head,
Clean the gratings using a soft, lint-free cloth and isopropyl or denatured alcohol.
Check to make sure the DC offset, amplitude, read head skew, and axis speed are within allowable limits.
If the error reoccurs, contact your Giddings & Lewis service representative.