The system sent a temperature compensation request but temperature compensation is currently not available. The message displayed includes a number in parentheses at the end that indicates the specific reason as follows:
-2 = No part thermistors are available
-1 = A previously reported error disabled temperature compensation.
0 = Temperature compensation is not properly configured.
1-28 = The indicated thermistor has failed.
For more information, see "TempComp Error Codes".
What you should do:
If possible, correct the identified problem.
If the error reoccurs, contact your Giddings & Lewis service representative.