This value determines the number of levels that will be used in the calibration process. PC-DMIS will divide the number of hits (DefNumHits) by the number of levels to determine how many hits will be taken at each level.
Entry Name: ProbeQualNumLevels
Entry Type: Whole Number specifying number of levels. The default value is 2.