This registry entry sets the quality level for sensor readings. A laser sensor returns a value for the quality of a reading. If this sensor value is below the QualityThreshold value, the software throws away those readings.
For example, peaks that fall below the QualityThreshold are considered invalid, and the software displays them with a measurement value of 0 (zero).
For measurement rates under 1kHz, set the QualityThreshold to 4.0 to prevent false readings.
For measurement rates of 1kHz and above, set the QualityThreshold to 1.5 to utilize the sensor's dynamic capabilities.
Entry Name: QualityThreshold
Entry Type: Whole Number that specifies a percentage. The default value is 25.