The Ref Temp values contain the reference temperature from which temperature compensation adjustments need to be applied.
The amount of correction to be applied is based
on multiplying the material coefficient by the difference between
the current and reference temperatures.
Amount of Correction = Material Coefficient x (Current Temp – Reference
Temp)
If the current temperature is the same as the reference temperature the net effect is that no thermal compensation adjustment is applied.
Entry Name: RefTemp(0-3)
Entry Type: Character string that specifies the degrees in either Celsius or Fahrenheit. The default value is 0.0.
Number Key
Value |
Option |
0 |
X Axis |
1 |
Y Axis |
2 |
Z Axis |
3 |
Part |