This registry entry determines the nominal method that will be used to gather nominal information.
Entry Name: ScanDialogNomMode
Entry Type: Whole number that specifies the nominal method number. The default value is 0.
Nominal method key:
Value |
Type |
Description |
0 |
MASTER |
PC-DMIS will consider the measured data obtained the first time the scan is being learned as nominal data. Subsequent executions of this scan will be compared with this measured data. |
1 |
FINDNOMS |
When the FIND NOMS option is selected, PC-DMIS will pierce the CAD model to find the closest location on a CAD surface to the measured point. It will then set the nominals to the location found on the CAD surface. |
2 |
NOMINALS |
PC-DMIS will use any measured data the first time it performs the scan as nominal data. This option allows you to relearn the scan without relearning the nominals. |