UseTraxWithSP600

The default 3D calibration algorithm for the SP600 has been changed to Trax. Of course, you are free to try the algorithm works best for your particular situation.

If using Trax calibration for the SP600, the effective tip size generated from the calibration will differ from the design value.

If using Trax calibration for non-SP600 analog probes on the Wetzlar machines, the design value of the tip size is used because tip size deviation is handled differently.

If using non-Trax calibration, the design value of the tip size is used.

Entry Name: UseTraxWithSP600

Entry Type: Whole number that specifies TRUE (1) or FALSE (0). The default value will be either 0 or 1 depending on your machine type and configuration. A value of -1 means the default value hasn't been set yet.