Qualification Check

Type of Operation area showing Qualification Check option

After you calibrate the unit, a qualification check is recommended, but not required. Performing a qualification check provides you with information regarding the general accuracy of the wrist calibration and future measurements. You can also use a qualification check to check the errors of new tips added to new probe files.

To perform a qualification check:

  1. Access the Probe Utilities dialog box by selecting the Insert | Hardware Definition | Probe menu option.

  2. On the Probe Utilities dialog box, from the Active Tip list, select the angles you wish to use for the qualification check. You should select probe angles both used and not used in the wrist calibration.

  3. Click the Measure button to open the Measure Probe dialog box.

  4. Provide the parameters you wish to use during the calibration check including selecting the appropriate calibration tool.

  5. From the Type of Operation area, select the Qualification Check option.

  6. Click the Measure button.

  7. Follow any on-screen instructions.