Move and Measure Parameters

AlwaysSendPrehit = False - When this parameter is set to True, the system forces the transmission of the prehit and scan retract distances, regardless of the EnableUserParams setting.

AlwaysSendScanSpeed = False - When this parameter is set to True, the system forces the transmission of the scan speed with every scan, regardless of the EnableUserParams setting.

EnableUserParams = True - When this parameter is set to True (default), the system silently transmits probing parameters to the machine as part of the PC-DMIS measurement routine. If this parameter is set to False, the system does not silently transmit any probing parameter setting to the machine. This means that probing is done with calibration parameters, enforcing technology-free measurements. Moving speeds are still transmitted.

SendSpeedChanges = True - When this parameter is set to True (default), the system silently transmits move speed changes to the machine. If this parameter is set to False, the system does not silently transmit move speed changes to the machine.