TempComp-200 TempComp not available

The system sent a temperature compensation request but temperature compensation is currently not available. The message displayed includes a number in parentheses at the end that indicates the specific reason as follows:

-2 = No part thermistors are available

-1 = A previously reported error disabled temperature compensation.

0 = Temperature compensation is not properly configured.

1-28 = The indicated thermistor has failed.

For more information, see "TempComp Error Codes".

What you should do:

  1. If possible, correct the identified problem.

  2. If the error reoccurs, contact your Giddings & Lewis service representative.