CWS.ini

Section

Key

Default Value

Type of Value

Description

COM

DEBUG

0

Integer

This value determines if enable debugging is on or off:

Off = 0

On = 1

COM

Connected

0

integer

This value determines if the application connects to the device or not:

No = 0

Yes = 1

COM

Port

7

integer

This value sets the serial port number for CWS.

COM

Baudrate

115200

integer

This value is the USB connection speed (CHRocodile requires 921600).

GRAPHIC

Zero

50.0

Double

This value is the offset position [%] in the measurement range.

CONFIG

SensorHeadNo

0

Integer

This value is the sensor head number to use at startup.

CONFIG

ForceSensorHeadNo

0

Integer

This value determines if you can set the sensor head from within the application or if the application only uses the value set in the SensorHeadNo registry key:

0 = This value allows you to set the sensor head from the application (for example, with TKJ-OMM).

1 = This value forces the application to always uses the sensor head number defined by the SensorHeadNo in the CWS.ini file. The application cannot set the sensor head number.

CONFIG

CHECKSINGLEVALUES

1

integer

This value determines if the validation of single point measurement raw result is on or off:

Validation Off = 0

Validation On = 1

Debug

FileTrace.System.Commands

0

integer

This value sets the log CWS system commands in HMV debug files on or off:

Off = 0

On = 1

In the case of the CHRocodile 2SHS, the COM settings for Port, Baudrate, Parity, Databits, Stopbits, and Handshake are not relevant since this controller requires an Ethernet connection (IP = 192.168.170.2, port = 7891).