AutoShutterTimeToFireOnFovEntry

This entry provides the delay (in milliseconds) that PC-DMIS waits before firing the point detection command for a Region of Interest (ROI) that was not entirely in the Field of View (FOV) but has just now entirely entered the FOV.

This entry is only evaluated when the AutoShutterEnable entry is set to 1 (TRUE). If the stage is not stable, then point detection is not reliable.

This value relates to the Movement detected settle in the Live View Setup dialog box. For information, see the "Setting up the Live View" topic in the PC-DMIS Vision documentation.

Entry Name: AutoShutterTimeToFireOnFovEntry

Entry Type: Whole number that specifies milliseconds. The default value is 3000.