EndAngle2

Use the DccEndAng2 entry value to effect this change rather than this value. This value is not currently used.

This value allows you to define the default value for the End Angle option that will be used when an Auto sphere is created. This is a user supplied angle in decimal degrees. This field will accept values from 0 to 90 (which represents 0° to 90°). If the ending angle is 0°, the last row will be a single hit at the top of the sphere.

Entry Name: EndAngle2

Entry Type: Real Number to specify the angle in degrees. The default value is NONE ( ).

Do not use an ending angle value of more than 90. The risk of the probe shank hitting the part increases when you use an angle greater than 90°.

Remember, when entering starting and ending angles, that the IJK vector always looks down the centerline, and that PC-DMIS measures DCC spheres in a counterclockwise direction.