ProbeQualVisionIlluminationCalGradientMaximumResolutionAdjustments_Lamp(0-5)

Sets the maximum number of resolution adjustments made during the illumination calibration. If the number of data points between the ProbeQualVisionIlluminationCalGradientMaximum_Lamp(0-5) and ProbeQualVisionIlluminationCalGradientMinimum_Lamp(0-5) entry values is less than ProbeQualVisionIlluminationCalGradientMinimumSamples_Lamp(0-5), the resolution or increment size of the calibration data is cut in half and the calibration data collection is repeated.

Entry Name: ProbeQualVisionIlluminationCalGradientMaximumResolutionAdjustments_Lamp(0-5)

Entry Type: Accepts integer values from 0 to 8 inclusive. The default value is 4.