This probe changer is used with the RDS rotating head. It is defined as ZEISSTC_RDS.
To calibrate this probe changer, take three hits. With the probe changer running along the X axis at the back of the machine and the user facing the probe changer, the hits are as follows for each slot:
On top of the small square at the back right of the slot (0,0,1)
On the right side of the small square (1,0,0)
On the back of the small square (0,1,0)
(A) - Hit 1 (Top)
(B) - Hit 2 (Right Side)
(C) - Hit 3 (Back)
The proper settings for the RDS probe changer are as follows:
[ZEISS]
TCDepthClearance = -47.360000
TCVerticalClearance = 151.5000
TCLengthClearance = 31.8000
Note that these values were used with older versions of PC-DMIS (for example, version 2.X).