Calibrating the LSPX3SF / HR-X (HR-X3-P-SF) Probe Changer

The HR-X3-P and HR-X3-P-SF style ports have considerably different physical constructions and have different dimensional offsets. You cannot mix-and-match these port styles.

The calibration process for the LSPX3SF / HR-X (HR-X3-P-SF) probe changer represents the steps that you would take to prepare to use any of the probe changers that PC-DMIS supports. The process discussed here is specific to the LSPX3SF / HR-X (HR-X3-P-SF) probe changer.

No inserts or extensions are used for any of the port positions.

Leitz LSPX3SF / HR-X (HR-X3-P-SF) rack

Leitz LSPX3SF / HR-X (HR-X3-P-SF) probe changer

The LSPX3SF / HR-X (HR-X3-P-SF) stylus changer DOES NOT need to be mounted on the machine table parallel to either the X or Y axis. However, it must be oriented at an angle which matches an available wrist angle for loading and unloading. For mounting instructions, consult the documentation that came with your probe changer.

To calibrate your probe changer:

Step 1 - Select the LSPX3SF / HR-X (HR-X3-P-SF) Probe Changer

Step 2 - Define the Mount Point

Step 3 - Define the Number of Ports

Step 4 - Prepare for Calibration

Step 5 - Take the First Manual Hit

Step 6 - Take the Second Manual Hit

Step 7 - Take Manual Hits on the Top of Each Port

Step 8 - Review the Calibration Results

Step 9 - Recalibrate Individual Ports