Step 4 - Prepare for Calibration

To start the calibration process for the TP200 Probe Changer:

  1. Select the Calibrate tab in the Probe Changer dialog box (Edit | Preferences | Probe Changer):

Probe Changer dialog box - Calibrate tab

Probe Changer dialog box - Calibrate tab

  1. Click the Calibrate button.

  2. The Active probe file and Active tip items default to the current measurement routine settings. If necessary, change these to the probe and tip that you intend to use for the probe changer calibration.

  3. Open the lids on ports 3 and 4 on each side of the divider (center post) by sliding them back and slightly to the side until they click into place.

Open port lids prior to calibration

Open port lids prior to calibration

In the next step, you take the first manual hit.