Step 4 - Prepare for Calibration

To begin the calibration process for the TP20 Probe Changer:

  1. Select the Calibrate tab in the Probe Changer dialog box (Edit | Preferences | Probe Changer):

Probe Changer dialog box - Calibrate tab

Probe Changer dialog box - Calibrate tab

  1. Click the Calibrate button.

  2. The Active probe file and Active tip items default to the current measurement routine settings. If needed, change these to the probe and tip that you intend to use for the probe changer calibration.

  3. Open the lids on each port by sliding them back until they click into place.

PC-DMIS Message

Open all lids and remove any probe from port 1.

When ready, please take a hit in the center of the silver front face between ports 1 and 2.

Prompt to open port lids prior to calibration

Open port lids prior to calibration

Open port lids prior to calibration

In the next step, you take the first manual hit.