This table details some basic PC-DMIS functions you can control with the Zeiss jog box:

PC-DMIS Function Jog Box
End F1
Execute Continue F2
Erase Hit F3 or Space key
Save Mast Auto Dim. F4
Store Move F5 or the button on the XY Joystick

If MappedJogKeys = True, then the following defaults are used:

WARNING: The UTIL_DELETE function can be dangerous on a jog box button.

This should not be used with an analog probe! If this is used with an analog probe, the probe will drop before the dialog box appears. For analog probes, you should only use #13!

When prompted to attach or release a probe (Attach Probe, Release Probe with OK, FOR MC-850, and FOR MC-850), you must use the button on the jog box assigned to the End key for the OK button on the dialog box.

Manual Scan Mode ON and Manual Scan Mode OFF must be assigned to jog box buttons on Vast systems with a probe changer. This is because these modes use during the Probe Change calibration.

The following button combinations communicate directly with the controller and are preferable over the assigned keys.

Picking Up the Probe

Enter (BELOW) + Shift (ABOVE) + F1

Releasing the Probe

Enter (BELOW) + Shift (ABOVE) + F2

The universal 3D probe head releases the probe immediately!

Correcting the Probe Position

Enter (BELOW) + Shift (ABOVE) + F3

The probe of the universal 3D probe head may drop down!

Weight Balance of Universal 3D Probe Head

Enter (BELOW) + Shift (ABOVE) + F4

JogBoxKey15

JogBoxKey15 = 3 - This sets JogBoxKey15 to the functionality of JogBoxKey3 (Insert MovePoint).

Jog Box Keys on Large Jog Boxes

The jog box keys on the large full alpha jog boxes are set up in the following fashion:

1 3 5 7 9 11 13 15 17 19

2 4 6 8 10 12 14 16 18 20

The button on top of the XY jog box is 21.

To make it easier, the key pressed and its current setting appear in the message area in the lower left-hand corner of the screen.