Using Auto-Inspect Mode

Auto-Inspect mode provides automated inspection of a sequence of points using a Leica tracker. This process is essentially the same as the typical point inspection process, except that the process may run unattended as the tracker automatically moves from one position to the next one.

This process is often used for deformation measurements or repeated stability studies over a long-time range. Each of the positions that are going to be auto inspected are typically equipped with a separate reflector.

For example, some typical cases for Auto-Inspect might include:

To use Auto-Inspect mode:

  1. Open or create a measurement routine.

  2. Insert a Manual/DCC mode command and set it to DCC.

  3. Select the Insert | Flow Control Command | On Error menu item to insert an On Error command.

On Error dialog box

  1. From the Error Type list, select the Reflector Not Found option.

  2. From the Error Mode section, select the Skip Command option.

  3. Insert points for each mounted reflector into your measurement routine. To do this:

  1. Aim the tracker at the reflector.

  2. To take a hit, on your keyboard, press Ctrl + H then press the End key.

  1. Execute the measurement routine.

In Execution mode, PC-DMIS automatically measures each of these points as follows:

  1. The Leica tracker points to the first point (position).

  2. The laser locks onto the positions if possible. If no reflector is there, or no reflector has been found with the current search settings, PC-DMIS continues to the next feature.

  3. If the laser locks onto the reflector, it measures the point.

  4. The process is repeated (steps 1 through 3) until PC-DMIS has measured or skipped all points.

For each point that was skipped, PC-DMIS displays the error message "Reflector not Found" along with the feature ID, and the feature's coordinate location. You can then use this information to take corrective action for the skipped points.