This entry was added in PC-DMIS version 4.3 MR2 and later. It lets you revert to the PC-DMIS version 4.2 and earlier behavior for legacy profile dimensions only. It has no effect on geometric tolerance commands.
If this entry is set to 1 (TRUE) and you select the Form only option from the Control options area of the legacy Profile dialog box, PC-DMIS includes the size in Profile calculations.
If this entry is set to 0 (FALSE) and you select the Form only option from the Control options area of the legacy Profile dialog box, PC-DMIS ignores size in Profile calculations.
Entry Name: UseSizeForProfileDimension
Entry Type: Whole number of 0 or 1. The default value is 1.