While the SP600 analog probe was in use, the Renishaw AC2 interface card was commanded to perform an A/D conversion but that conversion was not complete within 200 microseconds. This error indicates a critical failure of the interface card.
What you should do:
To clear the error, power the MP down.
Power the MP back up.
If the error reoccurs, contact your Giddings & Lewis service representative.