Using Simplified Temperature Compensation

Simplified temperature compensation is available for machines that are connected to Hexagon controllers. Currently, only stationary products, Leica Trackers and Romer Portable Arms  are supported.

You can access Simplified Temperature Compensation through the menu (Edit | Preferences | Temperature Compensation Setup).

The following product lines and machine types are not supported. They must use the original temperature compensation command.

If you use a controller with the Sheffield interface, the UseTemperatureCompensationV2 registry entry in the Option section must be False. If it is set to True, that registry entry may affect your measurement routine's move points in unexpected ways.

More:

Temperature Compensation Setup

Temperature Compensation

Variable for Part Temperature in Temperature Compensation Command

Automatic Insertion of Temperature Compensation Command

Temperature Compensation Dimensioning

Thermal Expansion Origin Command

Temperature Gage

Using Advanced Temperature Compensation