If the Chromatic White Light Sensor (CWS) is the active probe in the measurement routine, the Laser tab is visible.
When you use a CWS, be aware that important information is displayed on the available control box indicators.
The CWS control box typically has the following indicators:
Intensity Bar
The Intensity Bar displays the intensity of the measurement signal in a logarithmic scale. The intensity value is commonly shown in another display near the Intensity Bar. The display shows the relative units as a numeric value between 0 and 999. This is important information because if you're measuring a poorly reflecting surface, the intensity of the reflected light can be low. In this case, you need to reduce the measurement rate. Conversely, over-modulation of the sensor (intensity reading: 999, blinking) can cause measurement errors.
Distance Bar
The Distance Bar displays the current measurement value in a linear scale.
The measured distance appears in another display as a number in μm near the Distance Bar. This allows you to see where in the range the sensor is currently placed.
CWS Dark Reference Command
The "PassthruToController" command and dialog box are designed to send commands to the NC controller. It is used only by Hexagon Engineering.
For Hexagon Application Engineers and Technical Support, you can find detailed information on the PassthruToController command and dialog box on the SalesForce site here.
You can use the prefix "CWS", which represents the Precitec controller (CWS), and the token "#" to send commands to the Precitec controller.
For example, to take a dark reference, in the Edit window, type the command CWS#$DRK.
CWS# - Sends the command to the Precitec controller.
$DRK - Starts taking the dark reference.
All Precitec controller commands need to start with $.
If there is no prefix (CWS#), the PassthruToController command is sent to the NC controller.
The solution works for:
FDC-SLC
FDC-VisionBox
It DOES NOT work with Embedded controllers.