Using Protected Mode to Protect Measurement Routines from Edits

Enabling Protected Mode

To protect a measurement routine from unwanted edits, you can select the Edit | Protected Mode menu option to enable Protected Mode. This menu option appears on the Edit menu after you close all measurement routines.

When you select this menu option, the Set Protected Mode Password dialog box appears. It has two options to define a password to enable Protected Mode:

To define the password, type the same string of characters into the two boxes, and click OK. When you enable Protected Mode, a check mark appears next to the Protected Mode option on the Edit menu.

The Protected Mode password supersedes and overwrites any other password you have defined. This means that if you have enabled password protection for the Setup Options dialog box (Edit | Preferences |Setup) or for an .ipd (Inspection Plan Default) file, you need to use the password you defined when you enabled Protected Mode.

The PC-DMIS administrator can create marked sets, which allow the execution of a partial measurement routine. The operator can execute marked sets in Protected Mode.

Effects of Protected Mode

To protect your measurement routine, when you are in Protected Mode, you cannot do the following:

You can still do the following:

You cannot modify the measurement routine or reporting templates in any way.

Visual Indicators

The main visual indicators that PC-DMIS is in Protected Mode are these:

How to Disable Protected Mode

To disable Protected Mode, do the following:

  1. Close all open measurement routines.

  2. Select Edit | Protected Mode to open the Protected Mode Password dialog box.

  3. Enter your password.

  4. Click OK.

PC-DMIS then removes all of the above editing restrictions.

If you are a member of the Protect-Programmer group, you can only disable certified measurement routines. For more information about the Protect-Programmer group, see "Certify".