Calibrating the SCP600 Probe Changer

These topics describe how to define and calibrate the SP600 Probe Changer.

Inserts or extensions are not used for any of the port positions.

A Renishaw SCP600 Stylus Change Rack

The SCP600 Probe Changer MUST BE mounted on the machine table parallel to either the X or Y axis. The stylus must be greater than 40 mm so the puck and body of the probe clear the stylus wrench used to hold the port open. For mounting instructions, consult the documentation that came with your probe changer.

To calibrate your probe changer:

Step 1 - Select the SCP600 Probe Changer

Step 2 - Define the Mount Point

Step 3 - Define the Ports

Step 4 - Prepare for Calibration

Step 5 - Take the First Manual Hit

Step 6 - Take the Second Manual Hit

Step 7 - Review the Calibration Results